| Availability: | |
|---|---|
| Quantity: | |
LCKC-6
1)Open the door cover, hold the right hand of the centrifuge to press the button on the right, and lift the door coverto the left, until the support blockis supported.
2) check the left ventricle, remove the items except the rotor, and clean out the ventricle.
3) check whether the host, rotor, random tools and random files are complete and correct according to thepacking list.
4) confirmed the power supply voltage is consistent with the requirements voltage (see behind the machine outletof the power supply voltage, the use of power for single-phase three-wire ac 220 v, 50 hz power supply), the unit isequipped with the power cord plug in an external power receptacle.5) after completing the above, make sure the rotor is loose before starting the machinelf there is no abnormalphenomenon, it will start to run;
5) check the centrifuge brake switch in the "O" status, press the power switch marked symbol "I"end, connectthe power supply, choose work time 1 to 10 min, press the start key first, and make the centrifuge in running state,press stop button again, without exception, debugging success.If there is abnormal sound in the centrifuge, pleaseclose it immediately. If you find any problems with the electrical equipment, please let the professional maintenance personnel repair it.
content is empty!