Precautions for use
Make sure the power is on all thetime, once be turned off ,must bepreheated and self-checked again.
When the instrument was turned off for a long period of time,you can't test until 12 hours after power restored with thecalibration is finished.
The newly installed device should not beused until power-on for 48 hours, and needdaily calibration during the first week,afterwards, calibration should be done every 48 hours.
Ⅰ. Keep the instrument clean.
Ⅱ. Follow the manual strictly for operation.
Ⅲ. Frequently tidy the instrument and remove dust by softand non-fiber-releasing cloth.
Ⅳ. Please cover sample ports with caps enclosed for dust-proof if instrument will not be used for long time.
Ⅴ. Stop using if instrument does not work properly. Afterconfirming the problem occurrence, please contact thecustomer service in time.
In addition
The Carbon Dioxide filterneed to be replaced whenthe device be used every10,000 times or 2 years.
A spare part will be found inaccessory box.